Service

MHRA device registration

No device can be placed on the British market until it is on the MHRA register. A manufacturer outside the UK can only get there through a UK Responsible Person.

The requirement

Every medical device, in vitro diagnostic and custom-made device must be registered with the MHRA before it is placed on the market in Great Britain. Systems and procedure packs are not exempt.

The MHRA registers the manufacturer and its devices. A manufacturer established outside the UK cannot register itself: the application has to come from its UK Responsible Person. A manufacturer in Northern Ireland that has already registered for the Northern Ireland market does not need to register again for Great Britain.

What registration needs

The application is built from the detail of each device, so the quality of the information you send decides how smoothly it goes:

  • manufacturer name, address and contact details as they appear on the label
  • each device’s name, model or catalogue number and classification
  • the GMDN term for each device, which also determines how the MHRA groups it for fees
  • the conformity route, with the declaration of conformity and the certificates behind it
  • instructions for use and, where relevant, the UK importer’s details

We check all of it before a submission goes in. Gaps in a declaration of conformity, mismatched model numbers and missing certificates are the usual reasons an application stalls.

Fees

MHRA registration fees are payable by the manufacturer, not by us.

From 1 April 2026 the MHRA charges an annual fee of £300 per GMDN Level 2 category for the 2026/27 year, replacing the old one-off registration fee. Each category is charged once a year however many products sit inside it, registrations made mid-year are charged pro rata, and the annual amount is calculated on the categories in place on 31 March. There is also a statutory fee for registering in a new category, and some changes to an existing registration attract a fee. The yearly figure is set by the MHRA and changes.

Keeping a registration current

Registration is not a one-off task. Devices are added and withdrawn, categories change, fees fall due and the MHRA asks questions. Each of those passes through the UK Responsible Person.

We keep the registration aligned with what you actually sell, tell you in advance when the annual calculation is about to be made, and handle the correspondence that comes with it.

Certificates of Free Sale

Export customers and some overseas regulators ask for a UK Certificate of Free Sale, which confirms the devices are registered here. We apply to the MHRA on your behalf and pass the certificates to you. The MHRA charges £75 per order for up to ten certificates, and £10 for each additional certificate on the same order.

What we cannot do

We do not decide whether a device is registered, how long the MHRA takes, or what the MHRA concludes. We cannot approve, certify or authorise a device, and we do not provide UKCA conformity assessment support. Our job is to submit complete, accurate registrations and to keep them accurate afterwards.

Start with your device list

Send your device list with classifications and current certificates and we will tell you what the registration requires, including the number of GMDN categories the MHRA is likely to charge for.