Guide

UKRP and manufacturer responsibilities compared

Some duties move to the UK Responsible Person. Most do not. Which is which, including registration, labelling and vigilance reporting.

Published 23 November 2026Last reviewed 23 November 2026

Appointing a UK Responsible Person moves some duties and leaves others exactly where they were. Manufacturers who assume it moves all of them tend to discover the difference at the wrong moment.

The split

Obligation Manufacturer UK Responsible Person
Design and manufacture Yes No
Conformity assessment Yes Confirms it was carried out
Declaration of conformity Prepares and signs Holds a copy
Technical documentation Prepares and maintains Holds it, produces it for the MHRA
CE or UKCA marking Applies and maintains No
MHRA device registration Supplies the information Submits and maintains it
Registration fees Pays No
Post-market surveillance Yes No
Vigilance and incident reporting Yes Passes incident information to the manufacturer, co-operates with action
Complaints Handles Passes them to the manufacturer
MHRA correspondence Answers on substance Receives it and forwards it
Label identity Shows the UKRP’s name and address where UKCA marking is used Provides that detail

Registration belongs to the manufacturer

The MHRA registers a manufacturer and its devices. The UK Responsible Person is the only body permitted to make the application for a non-UK manufacturer, but the registration itself is the manufacturer’s, and so is the accuracy of what it contains. If a model number, a classification or a GMDN term is wrong, the correction starts with the manufacturer.

Vigilance is the usual misunderstanding

Reporting serious incidents and field safety corrective actions is the manufacturer’s duty, and the UK Responsible Person does not take it over. What the UKRP does is act as the route: complaints and incident information reaching the UK address are passed to the manufacturer, and the UKRP co-operates with any corrective action.

In practice the manufacturer still needs a vigilance process, a person who owns it and the ability to report to the MHRA within the required timescales.

Labelling changes with the mark

Where a device carries the UKCA mark, the UK Responsible Person’s name and registered address must appear on the device, its outer packaging or its instructions for use. A device carrying only a CE mark does not need them. Devices carrying both need them, because the UKCA marking route requires it.

That is a design change, not a label change made at the last minute, so it belongs in the plan before the first shipment.

Where the two overlap

Three areas need agreement rather than a division of duties:

  • Who the MHRA contacts first. The UKRP receives the request; the manufacturer answers the substance. Both need to know the process before a request arrives with a deadline.
  • Who holds current documentation. The UKRP must be able to produce it, so the manufacturer has to keep the copy held by the UKRP current.
  • What happens when something is wrong. The UKRP must report to the MHRA if a manufacturer acts contrary to its obligations. That is easier to accept when the expectation is written into the appointment.

A practical check before you appoint

Ask any prospective UKRP what they will do on the day the MHRA asks for a technical file, and who answers a vigilance question that arrives on a Friday. The answer tells you whether the appointment is a mailbox or a service.

Sources

This guide describes UK requirements in general terms. It is not advice on your own devices. See the regulatory disclaimer.